  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 88
A-Class Mark II 79
Step
Action Displayed Message Comment
I
Lower and latch the Head
Lift Lever.
Using the UP or the DOWN
Key, set the Gain Number
to 00 and then press
ENTER.
Record this reading as a
TOF Value for Gain
Number 00 in the table.
GAIN TRAN
*00 (0 -
31)
<yyy>
- Or, for reflective media:
GAIN REFL
*00 (0 – 31)
<yyy>
Where ‘yyy’ represents the
current sensor reading.
After ENTER is pressed, an
asterisk (*) appears to denote
selection.
Sample Calibration Table
Gain Number Label Value TOF Value Difference Value
00 252 248
01 250
02 248
… …
31 09
Step
Action Displayed Message Comment
J
Use the keys to increment
the Gain Number by one
and then press the ENTER
Key. Record the TOF Value.
Repeat this step for each
Gain Number.
GAIN TRAN
*01 (0 -
31)
<yyy>
- Or, for reflective media:
GAIN REFL
*01 (0 – 31)
<yyy>
Where ‘yyy’ represents the
current sensor reading.
Sample Calibration Table
Gain Number Label Value TOF Value Difference Value
00 252 248
01 250 245
02 248 234
… … …
31 09 14
(Continued)